Abstract

Optical measuring devices are widely used in industrial production and maintenance processes. With increasing integration, the accessibility of functional elements inside machines is limited for measuring devices. In order to provide a new measurement tool for these applications, the design of an endoscopic 3D sensor based on the structured light principle is presented. Application specific sensor heads are introduced, with diameters down to 8 mm attached to endoscopes with a length of 1 m. The sensor features a flexible pattern generator and a camera, allowing for dynamic reconfiguration of structured illumination patterns. Based on a brief comparison of existing light coding approaches, two different measuring modes for the sensor are described. The first measuring mode uses conventional phase-shift sequences in order to achieve highly accurate measurements under controlled conditions. The second measuring mode uses a new interleaved encoding approach for phase-shift patterns, which enables robust measurements even in the presence of relative motion between object and sensor. An example for this type of motion is the movement of the endoscope tip by hand. Using the proposed sequence design, the 3D acquisition rate is identical to the capture rate of the camera, which amounts to a maximum of 180 Hz in the current hardware configuration of the measuring endoscope. In combination with a registration and merging algorithm, consecutive measurements are automatically combined, enabling the measurement of large surfaces. Moreover, the approach is capable of estimating the relative location and orientation of the endoscope tip using odometry based on depth data. The performance of the sensor in both measuring modes is evaluated on the basis of measurements of reference objects. Furthermore, the results of a hand-held measurement of a reference standard are shown.

ASJC Scopus Sachgebiete

Zitieren

A 3D measuring endoscope for hand-guided operation. / Matthias, Steffen; Kästner, Markus; Reithmeier, Eduard.
in: Measurement Science and Technology, Jahrgang 29, Nr. 9, 094001, 01.08.2018.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Matthias S, Kästner M, Reithmeier E. A 3D measuring endoscope for hand-guided operation. Measurement Science and Technology. 2018 Aug 1;29(9):094001. doi: 10.15488/5158, 10.1088/1361-6501/aad1e1
Matthias, Steffen ; Kästner, Markus ; Reithmeier, Eduard. / A 3D measuring endoscope for hand-guided operation. in: Measurement Science and Technology. 2018 ; Jahrgang 29, Nr. 9.
Download
@article{2e19cf89162a4e5e8a86c0b76d70b9f9,
title = "A 3D measuring endoscope for hand-guided operation",
abstract = "Optical measuring devices are widely used in industrial production and maintenance processes. With increasing integration, the accessibility of functional elements inside machines is limited for measuring devices. In order to provide a new measurement tool for these applications, the design of an endoscopic 3D sensor based on the structured light principle is presented. Application specific sensor heads are introduced, with diameters down to 8 mm attached to endoscopes with a length of 1 m. The sensor features a flexible pattern generator and a camera, allowing for dynamic reconfiguration of structured illumination patterns. Based on a brief comparison of existing light coding approaches, two different measuring modes for the sensor are described. The first measuring mode uses conventional phase-shift sequences in order to achieve highly accurate measurements under controlled conditions. The second measuring mode uses a new interleaved encoding approach for phase-shift patterns, which enables robust measurements even in the presence of relative motion between object and sensor. An example for this type of motion is the movement of the endoscope tip by hand. Using the proposed sequence design, the 3D acquisition rate is identical to the capture rate of the camera, which amounts to a maximum of 180 Hz in the current hardware configuration of the measuring endoscope. In combination with a registration and merging algorithm, consecutive measurements are automatically combined, enabling the measurement of large surfaces. Moreover, the approach is capable of estimating the relative location and orientation of the endoscope tip using odometry based on depth data. The performance of the sensor in both measuring modes is evaluated on the basis of measurements of reference objects. Furthermore, the results of a hand-held measurement of a reference standard are shown.",
keywords = "3D endoscopy, motion, odometry, structured light",
author = "Steffen Matthias and Markus K{\"a}stner and Eduard Reithmeier",
note = "Funding information: The authors would like to thank the German Research Foundation (DFG) for funding the project B6 {\textquoteleft}Endoscopic geometry inspection{\textquoteright} within the Collaborative Research Center (CRC)/TR 73.",
year = "2018",
month = aug,
day = "1",
doi = "10.15488/5158",
language = "English",
volume = "29",
journal = "Measurement Science and Technology",
issn = "0957-0233",
publisher = "IOP Publishing Ltd.",
number = "9",

}

Download

TY - JOUR

T1 - A 3D measuring endoscope for hand-guided operation

AU - Matthias, Steffen

AU - Kästner, Markus

AU - Reithmeier, Eduard

N1 - Funding information: The authors would like to thank the German Research Foundation (DFG) for funding the project B6 ‘Endoscopic geometry inspection’ within the Collaborative Research Center (CRC)/TR 73.

PY - 2018/8/1

Y1 - 2018/8/1

N2 - Optical measuring devices are widely used in industrial production and maintenance processes. With increasing integration, the accessibility of functional elements inside machines is limited for measuring devices. In order to provide a new measurement tool for these applications, the design of an endoscopic 3D sensor based on the structured light principle is presented. Application specific sensor heads are introduced, with diameters down to 8 mm attached to endoscopes with a length of 1 m. The sensor features a flexible pattern generator and a camera, allowing for dynamic reconfiguration of structured illumination patterns. Based on a brief comparison of existing light coding approaches, two different measuring modes for the sensor are described. The first measuring mode uses conventional phase-shift sequences in order to achieve highly accurate measurements under controlled conditions. The second measuring mode uses a new interleaved encoding approach for phase-shift patterns, which enables robust measurements even in the presence of relative motion between object and sensor. An example for this type of motion is the movement of the endoscope tip by hand. Using the proposed sequence design, the 3D acquisition rate is identical to the capture rate of the camera, which amounts to a maximum of 180 Hz in the current hardware configuration of the measuring endoscope. In combination with a registration and merging algorithm, consecutive measurements are automatically combined, enabling the measurement of large surfaces. Moreover, the approach is capable of estimating the relative location and orientation of the endoscope tip using odometry based on depth data. The performance of the sensor in both measuring modes is evaluated on the basis of measurements of reference objects. Furthermore, the results of a hand-held measurement of a reference standard are shown.

AB - Optical measuring devices are widely used in industrial production and maintenance processes. With increasing integration, the accessibility of functional elements inside machines is limited for measuring devices. In order to provide a new measurement tool for these applications, the design of an endoscopic 3D sensor based on the structured light principle is presented. Application specific sensor heads are introduced, with diameters down to 8 mm attached to endoscopes with a length of 1 m. The sensor features a flexible pattern generator and a camera, allowing for dynamic reconfiguration of structured illumination patterns. Based on a brief comparison of existing light coding approaches, two different measuring modes for the sensor are described. The first measuring mode uses conventional phase-shift sequences in order to achieve highly accurate measurements under controlled conditions. The second measuring mode uses a new interleaved encoding approach for phase-shift patterns, which enables robust measurements even in the presence of relative motion between object and sensor. An example for this type of motion is the movement of the endoscope tip by hand. Using the proposed sequence design, the 3D acquisition rate is identical to the capture rate of the camera, which amounts to a maximum of 180 Hz in the current hardware configuration of the measuring endoscope. In combination with a registration and merging algorithm, consecutive measurements are automatically combined, enabling the measurement of large surfaces. Moreover, the approach is capable of estimating the relative location and orientation of the endoscope tip using odometry based on depth data. The performance of the sensor in both measuring modes is evaluated on the basis of measurements of reference objects. Furthermore, the results of a hand-held measurement of a reference standard are shown.

KW - 3D endoscopy

KW - motion

KW - odometry

KW - structured light

UR - http://www.scopus.com/inward/record.url?scp=85051679710&partnerID=8YFLogxK