Sensitivity vs Competing Proton Transfer Reactions: Addressing Key Parameters of Ion Chemistry in Ion Mobility Spectrometry

Research output: Contribution to journalArticleResearchpeer review

View graph of relations

Details

Original languageEnglish
Pages (from-to)1929-1939
Number of pages11
JournalJournal of the American Society for Mass Spectrometry
Volume36
Issue number9
Early online date1 Aug 2025
Publication statusPublished - 3 Sept 2025

Abstract

Ion mobility spectrometers (IMS) are widely used in various gas sensing applications due to their high sensitivity and rapid analysis times. However, in complex gas mixtures, reactions between the protonated target analyte and interfering species can lead to discrimination of analytes with low gas basicity, reducing sensitivity or even making detection impossible. Operating IMS at low pressure and high reduced electric field strengths has been shown to mitigate these competing ion–molecule reactions. Therefore, in this work, we present a kinetic model to evaluate the effect of key operating parameters on the ion suppression caused by competing ion–molecule reactions, guiding the instrumental design of IMS. The results demonstrate that measures to reduce competing ion–molecule reactions, such as lowering the operating pressure or reaction time, also reduce sensitivity due to fewer ion–neutral collisions. However, in scenarios with high concentrations of interferents, the reduced effect of competing ion–molecule reactions is critical for detecting target analytes with low gas basicity, thereby enhancing sensitivity under such conditions. Based on these findings, decreasing operating pressure and reaction time or increasing reduced electric field strength are the most promising strategies to minimize competing reactions and, in complex chemical backgrounds, increase sensitivity.

ASJC Scopus subject areas

Cite this

Sensitivity vs Competing Proton Transfer Reactions: Addressing Key Parameters of Ion Chemistry in Ion Mobility Spectrometry. / Schaefer, Christoph; Zimmermann, Stefan.
In: Journal of the American Society for Mass Spectrometry, Vol. 36, No. 9, 03.09.2025, p. 1929-1939.

Research output: Contribution to journalArticleResearchpeer review

Download
@article{d7f4fd7f456442b8b27a190f4f02d333,
title = "Sensitivity vs Competing Proton Transfer Reactions: Addressing Key Parameters of Ion Chemistry in Ion Mobility Spectrometry",
abstract = "Ion mobility spectrometers (IMS) are widely used in various gas sensing applications due to their high sensitivity and rapid analysis times. However, in complex gas mixtures, reactions between the protonated target analyte and interfering species can lead to discrimination of analytes with low gas basicity, reducing sensitivity or even making detection impossible. Operating IMS at low pressure and high reduced electric field strengths has been shown to mitigate these competing ion–molecule reactions. Therefore, in this work, we present a kinetic model to evaluate the effect of key operating parameters on the ion suppression caused by competing ion–molecule reactions, guiding the instrumental design of IMS. The results demonstrate that measures to reduce competing ion–molecule reactions, such as lowering the operating pressure or reaction time, also reduce sensitivity due to fewer ion–neutral collisions. However, in scenarios with high concentrations of interferents, the reduced effect of competing ion–molecule reactions is critical for detecting target analytes with low gas basicity, thereby enhancing sensitivity under such conditions. Based on these findings, decreasing operating pressure and reaction time or increasing reduced electric field strength are the most promising strategies to minimize competing reactions and, in complex chemical backgrounds, increase sensitivity.",
author = "Christoph Schaefer and Stefan Zimmermann",
note = "Publisher Copyright: {\textcopyright} 2025 The Authors. Published by American Chemical Society",
year = "2025",
month = sep,
day = "3",
doi = "10.1021/jasms.5c00161",
language = "English",
volume = "36",
pages = "1929--1939",
journal = "Journal of the American Society for Mass Spectrometry",
issn = "1044-0305",
publisher = "Elsevier Inc.",
number = "9",

}

Download

TY - JOUR

T1 - Sensitivity vs Competing Proton Transfer Reactions

T2 - Addressing Key Parameters of Ion Chemistry in Ion Mobility Spectrometry

AU - Schaefer, Christoph

AU - Zimmermann, Stefan

N1 - Publisher Copyright: © 2025 The Authors. Published by American Chemical Society

PY - 2025/9/3

Y1 - 2025/9/3

N2 - Ion mobility spectrometers (IMS) are widely used in various gas sensing applications due to their high sensitivity and rapid analysis times. However, in complex gas mixtures, reactions between the protonated target analyte and interfering species can lead to discrimination of analytes with low gas basicity, reducing sensitivity or even making detection impossible. Operating IMS at low pressure and high reduced electric field strengths has been shown to mitigate these competing ion–molecule reactions. Therefore, in this work, we present a kinetic model to evaluate the effect of key operating parameters on the ion suppression caused by competing ion–molecule reactions, guiding the instrumental design of IMS. The results demonstrate that measures to reduce competing ion–molecule reactions, such as lowering the operating pressure or reaction time, also reduce sensitivity due to fewer ion–neutral collisions. However, in scenarios with high concentrations of interferents, the reduced effect of competing ion–molecule reactions is critical for detecting target analytes with low gas basicity, thereby enhancing sensitivity under such conditions. Based on these findings, decreasing operating pressure and reaction time or increasing reduced electric field strength are the most promising strategies to minimize competing reactions and, in complex chemical backgrounds, increase sensitivity.

AB - Ion mobility spectrometers (IMS) are widely used in various gas sensing applications due to their high sensitivity and rapid analysis times. However, in complex gas mixtures, reactions between the protonated target analyte and interfering species can lead to discrimination of analytes with low gas basicity, reducing sensitivity or even making detection impossible. Operating IMS at low pressure and high reduced electric field strengths has been shown to mitigate these competing ion–molecule reactions. Therefore, in this work, we present a kinetic model to evaluate the effect of key operating parameters on the ion suppression caused by competing ion–molecule reactions, guiding the instrumental design of IMS. The results demonstrate that measures to reduce competing ion–molecule reactions, such as lowering the operating pressure or reaction time, also reduce sensitivity due to fewer ion–neutral collisions. However, in scenarios with high concentrations of interferents, the reduced effect of competing ion–molecule reactions is critical for detecting target analytes with low gas basicity, thereby enhancing sensitivity under such conditions. Based on these findings, decreasing operating pressure and reaction time or increasing reduced electric field strength are the most promising strategies to minimize competing reactions and, in complex chemical backgrounds, increase sensitivity.

UR - http://www.scopus.com/inward/record.url?scp=105015138101&partnerID=8YFLogxK

U2 - 10.1021/jasms.5c00161

DO - 10.1021/jasms.5c00161

M3 - Article

C2 - 40747918

AN - SCOPUS:105015138101

VL - 36

SP - 1929

EP - 1939

JO - Journal of the American Society for Mass Spectrometry

JF - Journal of the American Society for Mass Spectrometry

SN - 1044-0305

IS - 9

ER -