Abstract
Wear of the ultra-high molecular-weight polyethylene (UHMWPE) component in total knee arthroplasty contributes to implant failure. It is often detected late, when patients experience pain or instability. Early monitoring could enable timely intervention, preventing implant failure and joint degeneration. This study investigates the accuracy and precision (repeatability) of model-based wear measurement (MBWM), a novel technique that can estimate inlay thickness and wear radiographically. Six inlays were milled from non-crosslinked UHMWPE and imaged via X-ray in anteroposterior view at flexion angles 0°, 30°, and 60° on a phantom knee model. MBWM measurements were compared with reference values from a coordinate measurement machine. Three inlays were subjected to accelerated wear generation and similarly evaluated. MBWM estimated inlay thickness with medial and lateral accuracies of 0.13 ± 0.09 and 0.14 ± 0.09 mm, respectively, and linear wear with an accuracy of 0.07 ± 0.06 mm. Thickness measurements revealed significant lateral differences at 0° and 30° (0.22 ± 0.08 mm vs. 0.06 ± 0.06 mm, respectively; t-test, p = 0.0002). Precision was high, with average medial and lateral differences of − 0.01 ± 0.04 mm between double experiments. MBWM using plain radiographs presents a practical and promising approach for the clinical detection of implant wear.
Keywords
- In-vivo wear, Inlay wear, Minimum thickness, Model-based, Polyethylene wear, Standard radiograph
ASJC Scopus subject areas
Cite this
- Standard
- Harvard
- Apa
- Vancouver
- BibTeX
- RIS
In: Scientific reports, Vol. 14, No. 1, 18248, 12.2024.
Research output: Contribution to journal › Article › Research › peer review
}
TY - JOUR
T1 - Early monitoring of inlay wear after total knee arthroplasty on plain radiographs using model-based wear measurement
AU - Emonde, Crystal Kayaro
AU - Hurschler, Christof
AU - Breuer, André
AU - Eggers, Max Enno
AU - Wichmann, Marcel
AU - Ettinger, Max
AU - Denkena, Berend
N1 - Publisher Copyright: © The Author(s) 2024.
PY - 2024/8/6
Y1 - 2024/8/6
N2 - Wear of the ultra-high molecular-weight polyethylene (UHMWPE) component in total knee arthroplasty contributes to implant failure. It is often detected late, when patients experience pain or instability. Early monitoring could enable timely intervention, preventing implant failure and joint degeneration. This study investigates the accuracy and precision (repeatability) of model-based wear measurement (MBWM), a novel technique that can estimate inlay thickness and wear radiographically. Six inlays were milled from non-crosslinked UHMWPE and imaged via X-ray in anteroposterior view at flexion angles 0°, 30°, and 60° on a phantom knee model. MBWM measurements were compared with reference values from a coordinate measurement machine. Three inlays were subjected to accelerated wear generation and similarly evaluated. MBWM estimated inlay thickness with medial and lateral accuracies of 0.13 ± 0.09 and 0.14 ± 0.09 mm, respectively, and linear wear with an accuracy of 0.07 ± 0.06 mm. Thickness measurements revealed significant lateral differences at 0° and 30° (0.22 ± 0.08 mm vs. 0.06 ± 0.06 mm, respectively; t-test, p = 0.0002). Precision was high, with average medial and lateral differences of − 0.01 ± 0.04 mm between double experiments. MBWM using plain radiographs presents a practical and promising approach for the clinical detection of implant wear.
AB - Wear of the ultra-high molecular-weight polyethylene (UHMWPE) component in total knee arthroplasty contributes to implant failure. It is often detected late, when patients experience pain or instability. Early monitoring could enable timely intervention, preventing implant failure and joint degeneration. This study investigates the accuracy and precision (repeatability) of model-based wear measurement (MBWM), a novel technique that can estimate inlay thickness and wear radiographically. Six inlays were milled from non-crosslinked UHMWPE and imaged via X-ray in anteroposterior view at flexion angles 0°, 30°, and 60° on a phantom knee model. MBWM measurements were compared with reference values from a coordinate measurement machine. Three inlays were subjected to accelerated wear generation and similarly evaluated. MBWM estimated inlay thickness with medial and lateral accuracies of 0.13 ± 0.09 and 0.14 ± 0.09 mm, respectively, and linear wear with an accuracy of 0.07 ± 0.06 mm. Thickness measurements revealed significant lateral differences at 0° and 30° (0.22 ± 0.08 mm vs. 0.06 ± 0.06 mm, respectively; t-test, p = 0.0002). Precision was high, with average medial and lateral differences of − 0.01 ± 0.04 mm between double experiments. MBWM using plain radiographs presents a practical and promising approach for the clinical detection of implant wear.
KW - In-vivo wear
KW - Inlay wear
KW - Minimum thickness
KW - Model-based
KW - Polyethylene wear
KW - Standard radiograph