Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Autorschaft

Externe Organisationen

  • Medizinische Hochschule Hannover (MHH)
Forschungs-netzwerk anzeigen

Details

OriginalspracheEnglisch
Aufsatznummer1
FachzeitschriftBiomedical microdevices
Jahrgang28
Ausgabenummer1
PublikationsstatusVeröffentlicht - 29 Dez. 2025

Abstract

Roughly 5% of the population is affected by hearing loss. In case of a functional middle ear impairment with decreased transmission, different types of active middle ear implants may restore hearing. To drive the round window membrane (RWM) for example, bone anchored actuators or floating mass transducer (FMT) have been used in clinics. Bone anchored devices are usually too big for the round window niche and FMTs display low displacement and force at low frequencies. This paper focuses on the stimulation of the RWM by a stator-based actuator that combines small size and bone anchoring. The design features a rotational-symmetric balanced armature core with flexible membrane parts manufactured micro technologically. Membranes and assembled prototypes were evaluated on the bench in the range of 0.1 kHz - 10 kHz and prototypes tested in human temporal bones. We demonstrate the feasibility of a balanced armature actuator of appropriate size for RW placement. With 1.4 mm outer diameter and 4.9 mm length, the actuator fits the anatomical constraints. Acoustic stimulation of the RWM resulted in a maximum output of approximately 80 eq. dB SPL at 300 Hz increasing to > 100 eq. dB SPL above 5 kHz. The measured output is comparable to clinically used devices. The design lacked sufficient robustness leading to a low yield in manufacturing and failures in experiments, although the output level was insensitive to static force loading up to 30 mN. Future developments require an increased maximum output level and improved robustness.

ASJC Scopus Sachgebiete

Zitieren

Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load. / Prediger, Maren S; Müller, Eileen; Glukhovskoy, Anatoly et al.
in: Biomedical microdevices, Jahrgang 28, Nr. 1, 1, 29.12.2025.

Publikation: Beitrag in FachzeitschriftArtikelForschungPeer-Review

Download
@article{9e9fc953fad14b9ab487d82784c23431,
title = "Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load",
abstract = "Roughly 5% of the population is affected by hearing loss. In case of a functional middle ear impairment with decreased transmission, different types of active middle ear implants may restore hearing. To drive the round window membrane (RWM) for example, bone anchored actuators or floating mass transducer (FMT) have been used in clinics. Bone anchored devices are usually too big for the round window niche and FMTs display low displacement and force at low frequencies. This paper focuses on the stimulation of the RWM by a stator-based actuator that combines small size and bone anchoring. The design features a rotational-symmetric balanced armature core with flexible membrane parts manufactured micro technologically. Membranes and assembled prototypes were evaluated on the bench in the range of 0.1 kHz - 10 kHz and prototypes tested in human temporal bones. We demonstrate the feasibility of a balanced armature actuator of appropriate size for RW placement. With 1.4 mm outer diameter and 4.9 mm length, the actuator fits the anatomical constraints. Acoustic stimulation of the RWM resulted in a maximum output of approximately 80 eq. dB SPL at 300 Hz increasing to > 100 eq. dB SPL above 5 kHz. The measured output is comparable to clinically used devices. The design lacked sufficient robustness leading to a low yield in manufacturing and failures in experiments, although the output level was insensitive to static force loading up to 30 mN. Future developments require an increased maximum output level and improved robustness.",
keywords = "Humans, Round Window, Ear/physiology, Equipment Design, Acoustic Stimulation/instrumentation, Electromagnetic Phenomena, Ear, Inner/physiology, Round window stimulation, Balanced armature, Electromagnetic actuator, Round window actuator, Active middle ear implant",
author = "Prediger, {Maren S} and Eileen M{\"u}ller and Anatoly Glukhovskoy and Konrad Bethmann and Berta, {Andr{\'a}s Bern{\'a}t} and Wurz, {Marc C} and Hannes Maier",
note = "{\textcopyright} 2025. The Author(s).",
year = "2025",
month = dec,
day = "29",
doi = "10.1007/s10544-025-00766-x",
language = "English",
volume = "28",
journal = "Biomedical microdevices",
issn = "1387-2176",
publisher = "Springer",
number = "1",

}

Download

TY - JOUR

T1 - Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load

AU - Prediger, Maren S

AU - Müller, Eileen

AU - Glukhovskoy, Anatoly

AU - Bethmann, Konrad

AU - Berta, András Bernát

AU - Wurz, Marc C

AU - Maier, Hannes

N1 - © 2025. The Author(s).

PY - 2025/12/29

Y1 - 2025/12/29

N2 - Roughly 5% of the population is affected by hearing loss. In case of a functional middle ear impairment with decreased transmission, different types of active middle ear implants may restore hearing. To drive the round window membrane (RWM) for example, bone anchored actuators or floating mass transducer (FMT) have been used in clinics. Bone anchored devices are usually too big for the round window niche and FMTs display low displacement and force at low frequencies. This paper focuses on the stimulation of the RWM by a stator-based actuator that combines small size and bone anchoring. The design features a rotational-symmetric balanced armature core with flexible membrane parts manufactured micro technologically. Membranes and assembled prototypes were evaluated on the bench in the range of 0.1 kHz - 10 kHz and prototypes tested in human temporal bones. We demonstrate the feasibility of a balanced armature actuator of appropriate size for RW placement. With 1.4 mm outer diameter and 4.9 mm length, the actuator fits the anatomical constraints. Acoustic stimulation of the RWM resulted in a maximum output of approximately 80 eq. dB SPL at 300 Hz increasing to > 100 eq. dB SPL above 5 kHz. The measured output is comparable to clinically used devices. The design lacked sufficient robustness leading to a low yield in manufacturing and failures in experiments, although the output level was insensitive to static force loading up to 30 mN. Future developments require an increased maximum output level and improved robustness.

AB - Roughly 5% of the population is affected by hearing loss. In case of a functional middle ear impairment with decreased transmission, different types of active middle ear implants may restore hearing. To drive the round window membrane (RWM) for example, bone anchored actuators or floating mass transducer (FMT) have been used in clinics. Bone anchored devices are usually too big for the round window niche and FMTs display low displacement and force at low frequencies. This paper focuses on the stimulation of the RWM by a stator-based actuator that combines small size and bone anchoring. The design features a rotational-symmetric balanced armature core with flexible membrane parts manufactured micro technologically. Membranes and assembled prototypes were evaluated on the bench in the range of 0.1 kHz - 10 kHz and prototypes tested in human temporal bones. We demonstrate the feasibility of a balanced armature actuator of appropriate size for RW placement. With 1.4 mm outer diameter and 4.9 mm length, the actuator fits the anatomical constraints. Acoustic stimulation of the RWM resulted in a maximum output of approximately 80 eq. dB SPL at 300 Hz increasing to > 100 eq. dB SPL above 5 kHz. The measured output is comparable to clinically used devices. The design lacked sufficient robustness leading to a low yield in manufacturing and failures in experiments, although the output level was insensitive to static force loading up to 30 mN. Future developments require an increased maximum output level and improved robustness.

KW - Humans

KW - Round Window, Ear/physiology

KW - Equipment Design

KW - Acoustic Stimulation/instrumentation

KW - Electromagnetic Phenomena

KW - Ear, Inner/physiology

KW - Round window stimulation

KW - Balanced armature

KW - Electromagnetic actuator

KW - Round window actuator

KW - Active middle ear implant